Device Size
140 mm × 156mm × 94 mm (without Humidifier)
262 mm × 156mm × 94mm (with Humidifier)
Environment
conditions
Normal operation
Transport and storage
Temperature:5 ~ 35
-20 ~ +60
Humidity:10% ~ 93%,Non-condensing
10% ~ 93%,Non-condensing
Atmospheric Pressure:760 ~ 1060 hPa
760 ~ 1060 hPa
Note: The pressure value is given by 100 pa [hPa], 1 centimeter water column [cmH2O] = 0.98 hPa.
Production date
It can be acquired on the device label.
Operating life
The device can be used for five years if the users operate, maintain, and clean it according to the user manual. The operating life can be extended if the critical components are replaced.
Working mode
Continuous mode, the continuous work time is no less than 8 hours.
SD Card
The capacity is no less than 8G,it can record the users` treatment date.
Power supply requirement
Alternating voltage:100 ~ 240 V;
Frequency: 50 Hz / 60 Hz;
Input power:≤2 A
Security type
Class II ,Type BF Applied Part
Waterproof level
IP22 –Drip proof equipment
Pressure range
4 ~ 25 hPa(±0.5 hPa).
Pressure display accuracy
±0.5 hPa
Pressure adjustable range
See section 3.4
Operating mode
See section 3.4
Ramp time
The range of adjustment is 0 ~ 60 minutes
Noise
When the output pressure of the device is 10 hPa, its noise (A level of weighted sound pressure level) is not greater than 30 dB.
The user connector specification
φ22 mm cone joint
Filter specification
The most penetrating particle size is 10μm
The highest pressure about the patient connector
Single level normal state,20 hPa; Single fault state, not more than 30 hP
Bi-Level normal state,25 hPa; Single fault state, not more than 40 hP
Ventilatory capacity
The average flow rate of the user connection should be greater than the corresponding flow value in the following table when the working pressure is set as the values listed in the table below
Testing pressure(hPa)
4
9
15
20
25
Average flow rate of the user connection (L/min)
80
90
90
70
70
Humidifier
a) Temperature range of the humidifier:Adjustable from
0 to 5.
b)The gas temperature of the user connection should be ≤43.
Capacity for Humidifier water tank
>240ml
Tube
Length 1.8 m,Error:±0.18 m,Ø22mm.
Weight
Without humidifier: <1 kg;Humidifier included:<1.6 kg
Dynamic pressure stability
Pressure(hPa)
Respiratory rate 10BPM
Respiratory rate 15BPM
Respiratory rate 20BPM
4
±0.5
±0.55
±0.60
9
±0.80
±0.85
±0.85
15
±0.95
±0.95
±0.95
20
±0.95
±0.95
±1.0
25
±1.0
±1.0
±1.0
Standard Tube
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask - KingCare | KingCare.net
Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask

Manufacturer bmc cpap Machine For Night Sleep Snoring With Cpap Mask

WhatsApp Chat
Send Message